MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SOFSILK; SUTURE, NONABSORBABLE, SILK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SOFSILK; SUTURE, NONABSORBABLE, SILK Back to Search Results
Model Number GS-63-M
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a robotic bowel resection, the surgical technician was trying to pull the suture from the package on the sterile field when the needles consistently popped off.The suture was never removed.They then used another device to resolve the issue in order to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of five devices.Visual inspection noted that a retainer was received containing five suture.Two of the needles were engaged to the sutures and anchored to the foam of the retainer.It was observed that the retainer was melted and smashed.Functionally, the sutures could not be removed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFSILK
Common Device Name
SUTURE, NONABSORBABLE, SILK
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7869984
Report Number9612501-2018-01633
Device Sequence Number1720100
Product Code GAP
UDI-Device Identifier10884521085077
UDI-Public10884521085077
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K981128
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberGS-63-M
Device Catalogue NumberGS-63-M
Device Lot NumberD8C1204X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Initial Date Received by Manufacturer 08/23/2018
Supplement Date Received by Manufacturer01/29/2019
03/26/2019
Initial Report FDA Received Date09/12/2018
Supplement Report FDA Received Date02/25/2019
04/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-