(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.As there were no problems advancing the unspecified 3.5x24 mm stent delivery system over the guide wire after the unspecified 2.0x15 mm balloon dilatation catheter was used, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely interaction with the balloon dilatation catheter resulted in the reported difficulty to position and the reported difficulty to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|