Model Number CI-1400-02H
Device Problem Electrical Shorting (2926)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2017
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing open and shorted electrodes and decreased performance.Revision surgery is scheduled.
 
Manufacturer Narrative
The recipient was reportedly reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed near the array prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed multiple broken electrode wires at the fantail.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed the electrical tests performed.The device passed the mechanical tests performed.The scanning electron microscopy analysis of the electrode revealed evidence of fatigue breaks on multiple wires at the fantail region.The photographic imaging inspection and the scanning electron microscopy analysis revealed evidence of fatigue breaks on multiple electrode wires at the fantail region.These breaks can be associated with the decrease in performance and the impedance failures reported.A corrective action was implemented.This is the final report.
 
Manufacturer Narrative
The external visual inspection revealed the electrode was cut prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed broken electrode wires near the fantail.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed the electrical tests performed.This is an interim report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
delilah garcia
28515 westinghouse place
valencia, CA 91355
MDR Report Key7870286
MDR Text Key120307133
Report Number3006556115-2018-00431
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/30/2013
Device Model NumberCI-1400-02H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2019
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age10 YR
-
-