| Model Number |
LS1520 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the tip of the device would not open properly after 3 firings.It occurred after 3-5 mins of being used.A second device was required.There was no medical intervention required.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The investigation found the jaws opened and closed properly when the latch was retracted and released.The device was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|