| Catalog Number |
21346 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer in the united states reported a misidentification of an acinetobacter lwoffii cap survey sample (sputum), as moraxella catarrhalis (92%) in association with the vitek® 2 nh test kit.The nh card result had low discrimination moraxella catarrhalis/campylobacter coli with contraindicating biopatterns, so the customer tested the isolate on the vitek gn card and obtained the result of moraxella group.A biomérieux internal investigation will be initiated.
|
| |
|
Additional Manufacturer Narrative
|
|
A us customer reported a misidentification of an acinetobacter lwoffii cap survey strain (d-09) as moraxella group with the vitek® 2 nh test kit.The nh card knowledge base does not include a.Lwoffii as a claimed species so either an unidentified result or misidentification may occur.The following limitation is contained in the product labeling: testing of unclaimed species may result in an unidentified result or a misidentification.
|
| |
|