| Catalog Number |
520A1300US |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation in progress.
|
| |
|
Event or Problem Description
|
|
It was reported that the pneupac ventilators vr1 air mix failed to stay in auto mode.No death or serious injury was reported in connection with this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: a smiths medical, vr1 ventilator was received in good condition.Performed initial check of momentary valve/auto switch.Found that switch would not stay in auto mode position, it would rotate back to manual position.This confirmed customer reported reason.Replaced momentary switch part number 520a1311.Tested to 504-2046.Device passed all functional tests.The customer reported condition was confirmed.Dhr review summary: previous service record is attached and review deemed it as unrelated to this complaint.Allegation confirmed: yes.Problem source details: possible normal wear and tear.
|
| |
|