Details of the complaint on (b)(6) 2018, customer at (b)(6) hospital reported the bedside monitor (bsm-2354a sn: (b)(6) had issues with the screen going in and out.The screen would be working but then would fade out and go to a white screen.If he taps on the screen, it will flicker but will still go to a white screen.Service requested evaluation: service performed unit was cleaned and under evaluation.We also visual inspected the labels that matched with the unit.The reported problem "this bsm has an issues with the screen going in and out" was duplicated.We powered on and tapped the on the screen, the lcd was faded out.We are unable to repair this bsm due to the lcd was not available at this time.Exchange was processed.Investigation result the bsm warranty began 07/07/2006.Issue was reported after 12 years at customer facility.Per nkc dhr (attached), the unit has had no history of ncmr, deviation, or capa during manufacturing of the device.The device has not been refurbished and there were no discrepancies or unusual findings before device release that might relate to the reported issue.Review of device sap history found previously reported issues with the unit: (1) 300092568 reported 08/04/2017 in which unit was reported to have damage.Parts were requested for customer to perform repairs.(2) 300098328 reported 09/22/2017 in which unit's screen was reported to dim intermittently.Nka evaluation was unable to duplicate the reported issue.A precautionary replacement of the lcd was recommended however due to lack of customer response, the unit was returned unrepaired.Issue with lcd display was reported shortly after device damage, however whether the damage contributed to lcd failure could not be confirmed.The bsm-2354 model was discontinued on 08/03/2011 in mbg-110308a due to the introduction of bsm-6000 series monitor.Support for the unit continued for a minimum period of seven years.The issue was reported well beyond the warranty period and product discontinuance, and at the end of promised support period.Due to a lack of parts available, the lcd could not be replaced.The root cause is determined to be continued use of the unit past its expected serviceable period.This issue is not suspected to be caused by deficient design.Additional information: b4.Date of this report.F6.Date user facility/importer became aware of the event.F7.Type of report.F11.Date report sent to fda.F13.Date report sent to manufacturer.G4.Date received by manufacturer.G7.Type of report.H2.If follow-up, what type? additional information.H3.Device evaluated by manufacturer.H6.Event problem and evaluation codes.H10.Additional manufacturer narrative.
|