| Model Number |
25212 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned to masimo for evaluation.When the investigation is complete a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported that the sphb readings are incorrect.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.During evaluation the device passed all visual and functional testing.The unit was found to visually and audibly alarm during alert conditions.During simulation testing, the device was found to provide accurate measurements.No performance issues were identified.The reported issue was unable to be confirmed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the sphb readings are incorrect.No patient impact or consequences were reported.
|
| |
|