| Model Number |
98814-65 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Unless further information is obtained, this issue is considered closed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
An extended investigation has been conducted, which involved a manufacturing review (including 5 years of device history records (dhrs), previous complaint and corrective and preventive action (capa) investigations conducted for blank screen issues, process failure mode effects analyses (pfmeas), design controls, and design failure mode effects analyses (dfmeas), risk management reports, risk evaluations, and label copy.The investigation did not identify any indication that the product did not meet specification.If the product is returned, an investigation will be performed.
|
| |
|
Event or Problem Description
|
|
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Search Alerts/Recalls
|