Catalog Number 124110527
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 08/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Concomitant medical products: total hip ball 28mm +5 hip femoral head; duraloc const liner 50x28 hip acetabular liner with duraloc dynamic lock ring 50mm hip acetabular liner locking ring; unknown hip acetabular cup; unknown hip femoral stem.
 
Event or Problem Description
It was reported that the patient has a right total hip with aml stem and duraloc cup.The patient had a constrained liner implanted in april and has dislocated.The ring to the liner was over the neck of the stem.The surgeon decided to remove the cup and implant another with better version.The stem was retained and is well fixed.Doi: (b)(6) 2018; dor: (b)(6) 2018; right hip.
 
Additional Manufacturer Narrative
Pc-(b)(4).No device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch , a follow-up medwatch will be filed as appropriate. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURALOC CONST LINER 50X28
Common Device Name
DURALOC IMPLANT : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0998
MDR Report Key7872392
Report Number1818910-2018-69257
Device Sequence Number2050279
Product Code KWZ
UDI-Device Identifier10603295018643
UDI-Public10603295018643
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P960054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number124110527
Device Lot Number548227
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/19/2018
Supplement Date Received by Manufacturer10/09/2018
Initial Report FDA Received Date09/13/2018
Supplement Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient Weight52
-
-