MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problems Computer Software Problem (1112); Failure to Obtain Sample (2533)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/22/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.No parts have been returned to the manufacturer for analysis.
 
Event Description
Medtronic received information regarding a navigation device.It was reported that system was displaying the trace pattern on the back of the head.Registration attempted four times without resolution.An alternate navigation system was utilized to complete the procedure.There was no impact to patient.There was a reported delay to the procedure of less than 1 hour due to this issue.
 
Manufacturer Narrative
Correction: correcting fdad code to proper value.A software analysis was initiated to determine the probable cause of the issue through review of the reported issue.By reviewing the case detail reported all the trace patterns were composed of a large amount of forehead points with very little unique anatomy such as nose.Analysis found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
koko tyler
navigation customer quality
826 coal creek circle
louisville, CO 80027
7208903200
MDR Report Key7872694
MDR Text Key120153524
Report Number1723170-2018-04626
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/22/2018
Initial Date FDA Received09/13/2018
Supplement Dates Manufacturer Received09/20/2018
Supplement Dates FDA Received09/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-