The returned valve was patent.It met the requirements for reflux, pressure-flow, preimplantation, and leak testing.Therefore, the conditions of the complaint could not be duplicated by laboratory personnel.The instructions for use that accompanies the device instructs that when performing the patency check: place the inlet connector of the valve into sterile isotonic saline; depress the valve dome; place a finger over the opening at the end of the outlet connector; release the depressed dome.If fluid enters the reservoir, the inlet connector and flow control membrane valve are patent; remove finger from the outlet connector opening; depress dome.If fluid flows out of the outlet connector, the valve is patent.Note: it may be necessary to depress the valve dome more than once to initiate flow, especially if the ventricular catheter has been attached prior to patency testing.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|