Model Number 27303
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was to undergo lumbar drainage due to intracranial hemorrhagic occupying inflammatory lesions.When the package was opened during the operation, it was found that the catheter was distorted.The catheter was not used and was replaced.There was no injury to the patient.
 
Additional Manufacturer Narrative
The returned catheter met the requirements for leak test.However, the catheter was noted to be distorted/kinked.It is not known how or when this damage occurred.There was no apparent damage to the packaging that the catheter and luerlock cap were originally packaged in, and subsequently returned in.Neither component were apparently used.The other components of the kit were returned unopened.No other unusual characteristics were noted.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTERNAL DRAINAGE AND MONITORING, LUMBAR DRAINAGE KIT
Common Device Name
DEVICE, MONITORING, INTRACRANIAL PRESSURE
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7873013
Report Number2021898-2018-00461
Device Sequence Number11154895
Product Code GWM
UDI-Device Identifier00673978246048
UDI-Public00673978246048
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K802100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Model Number27303
Device Catalogue Number27303
Device Lot Number212595875
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2018
Supplement Date Received by Manufacturer11/30/2018
Initial Report FDA Received Date09/13/2018
Supplement Report FDA Received Date12/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
-
-