Catalog Number 1012631-29
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation was unable to determine a conclusive cause for the reported inflation issue.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the omnilink elite stent delivery system (sds) was advanced to the common iliac, moderately calcified lesion without issue.Balloon inflation was attempted, however, the balloon failed to inflate.The sds with the undeployed stent was removed without issue.The stent remained on the delivery system with no dislodgment and no movement noted.The device was discarded and another omnilink was used in replacement.There was no adverse patient effect and there was no clinically significant delay.No additional information was provided regarding this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7873079
Report Number2024168-2018-07146
Device Sequence Number10819845
Product Code NIO
UDI-Device Identifier08717648178504
UDI-Public08717648178504
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Catalogue Number1012631-29
Device Lot Number8052441
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2018
Initial Report FDA Received Date09/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient Weight79
-
-