| Model Number |
N/A |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device has been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
|
| |
|
Event or Problem Description
|
|
During product evaluation of the returned precision device there was resistance that was felt when inserting the guidewire into the needle.The patient that was involved with the reported device was in stable condition.The procedure outcome was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the completed investigation results.One 21g needle and 0.021" guidewire were returned for evaluation.No sheath or dilator were returned.The returned needle and guidewire were subjected to visual analysis.There were no gross anomalies noted with the returned guidewire.The needle was examined under microscopy.At the cannula hub junction, there was flash.Dimensional testing was performed by measuring the inner lumen of the needle hub junction and the outer diameter of the guidewire.The inner diameter of the needle was then measured and confirmed to meet manufacturer specifications.The outer diameter of the guidewire was measured and confirmed to meet manufacturer specifications.Functional testing was then performed by inserting the guidewire through the needle.The guidewire passed through the needle.However, there was some resistance felt when initially passing the guidewire through the needle.There is no evidence that this event was related to a device defect or malfunction.Based on the investigation results the device met dimensional specifications.However, the exact cause of the reported event cannot be definitively determined.
|
| |
|
Search Alerts/Recalls
|