• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. PROMOTE PLUS CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. PROMOTE PLUS CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1233-40Q
Medical Device Problem Codes Premature Discharge of Battery (1057); Failure to Interrogate (1332)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/28/2018
Type of Reportable Event Serious Injury
Event or Problem Description
During a follow-up in clinic, there was an unexpected decrease in remaining longevity of the device.Premature battery depletion was suspected.Device replacement is anticipated and the patient was stable with no consequences.
 
Additional Manufacturer Narrative
The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on 11 october 2016.
 
Event or Problem Description
Additional information received indicated that the device could not be interrogated.The device was explanted and the patient was stable following the procedure.
 
Additional Manufacturer Narrative
Premature battery depletion was confirmed by analysis.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE PLUS CRT-D
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
MDR Report Key7873610
Report Number2938836-2018-10599
Device Sequence Number1190879
Product Code LWS
UDI-Device Identifier05414734503693
UDI-Public05414734503693
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2013
Device Model NumberCD1233-40Q
Device Lot Number3610270
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer10/23/2018
11/14/2018
01/22/2019
Initial Report FDA Received Date09/13/2018
Supplement Report FDA Received Date10/25/2018
11/16/2018
01/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0115-2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-