• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION PROPAQ MD DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION PROPAQ MD DEFIBRILLATOR Back to Search Results
Model Number PROPAQ MD
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that during biomed testing, the device displayed a "short detected" message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling and ecg monitoring stress testing on a simulator without duplicating the report.The device was recertified and returned to the customer.The multi-function cable used at the time of the event was not returned as part of this investigation.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPAQ MD DEFIBRILLATOR
Common Device Name
PROPAQ MD
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key7874044
Report Number1220908-2018-02536
Device Sequence Number16698628
Product Code DRT
UDI-Device Identifier00847946017651
UDI-Public00847946017651
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K100654
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberPROPAQ MD
Device Catalogue NumberPROPAQ MD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/23/2018
Supplement Date Received by Manufacturer08/23/2018
Initial Report FDA Received Date09/13/2018
Supplement Report FDA Received Date09/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-