WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11MM/125 DEG TI CANN TROCH FIXATION NAIL 170MM-STERILE; ROD,FIXATION,INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
456.317S |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Not Applicable (3189); No Code Available (3191)
|
| Date of Event |
08/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional product code hwc.Device malfunctioned intra-operatively and was not implanted / explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, during a trochanteric fixation nail-advanced (tfna) procedure, an 11mm/125 degree titanium cannulated trochanteric fixation nail (170mm) was inadvertently selected for implantation.The surgeon had difficulty when attempting to implant the nail.The correct, larger, nail (11 mm/130 degree titanium cannulated tfn-a) was then selected and switched onto the aiming arm for successful implantation.There was a 20-minute surgical delay reported.The procedure was successfully completed.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: a device history record (dhr) review was conducted.Manufacturing location: monument.Manufacturing date: 01-may-2013.Expiration date: 31-mar-2022.Part number: 456.317s, 11mm/125 deg ti cann troch fixation nail 170mm-sterile lot number: 7370479 (sterile).Lot quantity: 6.Work order traveler met all inspection acceptance criteria.Inspection sheet, in-process acceptance sheet, met all inspection acceptance criteria.Inspection sheet, inspect dimensional, met all inspection acceptance criteria.Inspection sheet, final inspection, met all inspection acceptance criteria.Packaging label log, lppf, lmd/lpf rev h was reviewed and determined to be conforming.Packaging bom was reviewed and all components used met current defined requirements.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component parts reviewed: component parts were not reviewed because the reported complaint condition was relevant to an incorrect nail being supplied and, therefore, a review of the component dhrs would not be relevant to the reported complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation, as part was discarded by the facility.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: (b)(6).: due to the intra-operative events, the device was not successfully implanted.An alternate device was used to complete procedural step.As such, implant/explant dates are not applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Further it was reported that there was no patient consequences and patient outcome was fine.
|
| |
|