| Catalog Number |
11-1020 |
| Medical Device Problem Codes |
Fitting Problem (2183); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation in progress.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
|
| |
|
Event or Problem Description
|
|
It was reported that the mdi canister on the portex oxygen/aerosol ace kit could not be placed in the canister port/dispenser.
|
| |
|
Additional Manufacturer Narrative
|
|
Two breathing portex units received.The end of the conical unit was confirmed to be distorted, which prevented user from assembling.Upon inspection of the units, it was discovered that there were markings indicating the units were compressed.The markings were noted to be on both the spherical and conical ends indicating the distortion happened after molding and welding, and it was confirmed the t distorted end of conical units would not fit in the welding jig.The customer complaint ws confirmed, however a thorough investigation of the manufacturing process could not duplicate the fault found on the samples.The process investigation and complaints do not signify the fault is trending, therefore no action will be taken at this time.
|
| |
|