| Catalog Number |
3504310M |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
07/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent a primary breast reconstruction with mentor tissue expander exp rect remote 550cc implant which the left side deflated after implantation.In the morning (b)(6), the implant was empty.Patient had been sleeping on her left side where the implant was placed.As a result patient had the device removed and replaced with catalog number 3504201m on (b)(6) 2018.The patient did not experience any accompanying symptoms, was not hospitalized, and has since fully recovered.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: upon receipt by mentor, the device returned without fluid.No foreign material was observed within the device or on the shell surface.Product evaluation team discovered a 2 cm rent at the patch junction.Microscopic examination of the edges of the rent gave no indications as to cause.No other anomalies were discovered.The complaint was confirmed since a rupture was found on the device.However, at this point it is not possible to determine the root cause of such damage.All the implants are 100% inspected before leaving the facility.There is no evidence that the issue is related with manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|