| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Missing Value Reason (3192)
|
| Date of Event |
08/15/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Lens was returned dry, in a specimen container.Visual inspection found a torn haptic and residue on the lens surface.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm, micl13.2, -14.0 diopter, implantable collamer lens into the patients right eye (od) on (b)(6) 2018.The lens was exchanged on (b)(6) 2018 with a shorter length lens due to pupil block with elevated iop (intraocular pressure), too large lens with ac shallowing, high vaulting and narrowing of angle.This exchange resolved the problem.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|