• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 2.5MM HEX LINEAR DRIVER SHAFT; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 2.5MM HEX LINEAR DRIVER SHAFT; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE Back to Search Results
Catalog Number 71175098
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Information (3190)
Date of Event 08/24/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that in the proximal tibia the doctor used an evos 8h lateral proximal tibia plate.When doctor inserted the most distal screw on power (percutaneous), then final tightened on hand the screw stripped.
 
Additional Manufacturer Narrative
The associated complaint devices were returned for evaluation.A visual assessment of the devices confirmed the stated failure.The hex end of the driver shaft was rounded and the prongs on the capture sleeve were damaged/bent.The device exhibits signs of significate wear/usage.A review of complaint history did not reveal any additional complaints for the listed batch for the same failure mode.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No additional actions are being taken at this time; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.5MM HEX LINEAR DRIVER SHAFT
Common Device Name
INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7875161
Report Number1020279-2018-01785
Device Sequence Number16191040
Product Code MDM
UDI-Device Identifier00885556651339
UDI-Public00885556651339
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71175098
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/24/2018
Supplement Date Received by Manufacturer08/24/2018
Initial Report FDA Received Date09/13/2018
Supplement Report FDA Received Date11/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-