MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22306D
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that during a therapeutic transurethral resection of the prostate (turp) procedure, the loop wire at the distal end of the hf resection electrode broke off and fell inside the patient.However, no fragment remained inside the patient since it was reportedly retrieved.The intended procedure was successfully completed with the same set of equipment and there was no report about an adverse event or patient injury.
 
Additional Manufacturer Narrative
The suspect medical device was not returned to olympus for evaluation/investigation since it was reportedly discarded by the user facility.Therefore, the exact cause of the user's experience and the reported phenomenon could not be conclusively determined and is being judged as unknown.However, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°
Common Device Name
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
MDR Report Key7876061
Report Number9610773-2018-00085
Device Sequence Number186672
Product Code FAS
UDI-Device Identifier14042761051672
UDI-Public14042761051672
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K100275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/23/2023
Device Model NumberWA22306D
Device Catalogue NumberWA22306D
Device Lot Number1000022498
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/07/2018
Supplement Date Received by Manufacturer11/05/2018
Initial Report FDA Received Date09/14/2018
Supplement Report FDA Received Date11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-