| Model Number |
SPL-PDBX376 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
08/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to cybersonic for evaluation.A review of similar reported complaints indicate that this reported phenomenon is likely attributed to excessive force applied on the device and/ or excessive torqing during the procedure.It is not clear how the additional hole was caused practically.
|
| |
|
Event or Problem Description
|
|
The user facility reported that during a lithotripsy procedure the lithotripsy probe broke off six inches from the seating thread.An extra hole was made in an attempt to locate the broken piece.
|
| |
|