• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK-HINGE FIXED TIBIAL INSERT SIZE 5/12MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK-HINGE FIXED TIBIAL INSERT SIZE 5/12MM Back to Search Results
Catalog Number 02.09.0512H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 13 september 2018.Lot 120031ar: (b)(4) items manufactured and released on 02 april 2014.Expiration date: 2017-04-30 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
We were informed on 17 august 2018 that a patient came in complaining of pain.The surgeon believes the poly is loose.The surgeon plans to revise on (b)(6) 2018.More information to follow the case.Update 21 august 2018: on (b)(6) 2018, during surgery, the surgeon determined that the poly fixation screw broke.The surgeon left everything as is, the screw was not retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-HINGE FIXED TIBIAL INSERT SIZE 5/12MM
Common Device Name
FIXED TIBIAL INSERT
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
MDR Report Key7876591
Report Number3005180920-2018-00689
Device Sequence Number192300
Product Code KRO
UDI-Device Identifier07630030825262
UDI-Public07630030825262
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K130299
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2017
Device Catalogue Number02.09.0512H
Device Lot Number120031AR
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2018
Initial Report FDA Received Date09/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-