Medical Device Problem Codes Premature Discharge of Battery (1057); Moisture Damage (1405)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a power (battery life w/ moisture) issue.The reporter stated there was moisture in the battery compartment and denied any damage to the battery cap or battery compartment.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 28-sep-2018 with the following findings: evidence of short battery life was illustrated in the pump history with ¿cs128¿ (replace battery alarm) occurring due drastic voltage drop on (b)(6)-2018.The battery compartment was cracked at the threads on the side and separately below the grip pad.Evidence of moisture corrosion was observed in the battery compartment.A leak test failed due a battery compartment leak.The ¿ez-prime¿ steps were performed correctly; all current readings were within specifications.The pump¿s cover was removed, no further moisture ingress or any intermittent condition was found to the power printed circuit board or to the continuous glucose monitor module.Investigation was unable to duplicate the ¿short battery life¿ complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key7876867
Report Number2531779-2018-16837
Device Sequence Number16659331
Product Code OYC
UDI-Device Identifier10840406100181
UDI-Public10840406100181
Combination Product (Y/N)Y
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2018
Was the Report Sent to FDA? Yes
Device Age24 MO
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer09/04/2018
Initial Report FDA Received Date09/14/2018
Supplement Report FDA Received Date10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age17 YR
Patient Weight87
-
-