This event has been recorded in (b)(6).Three (3) units were returned to our facility for testing.In two (2) of them, visual inspection failed and flow rate testing could not be performed because its capillary element was partially blocked (probably because of drug crystallization).The other returned unit passed visual inspections and standardized flow rate test.In addition, five (5) units from the same manufacturing lot were retrieved from retained samples and successfully passed visual inspection and flow rate testing.Therefore, as it was not possible to determine the flow rate of two (2) of the returned units, and since both the remaining returned unit and the units from the archives of samples were free of defects, it was not possible to determine the root cause of the defect reported.
|
|
In 5-fu medication scheduled for 46 hours, infusion time finished much earlier than expected.Patient received 5-fu 3200mg and it was expected to finish around noon.However, the infusion finished 10 hours early (the accuracy of the infusion time is +/- 10%).Patient experienced adverse effects of face and tongue swelling that lasted 2 days, but denied nausea / vomiting / diarrhea.
|