| Catalog Number |
PXT231416 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Atrial Perforation (2511)
|
| Date of Event |
08/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2010, this patient underwent endovascular repair of an infrarenal abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis.It was reported that the patient had a significant aortic neck angle approximately 2-3 cm below the lowest renal artery.On (b)(6) 2018, during follow up angiography, it was suspected that the aortic neck of the trunk-ipsilateral leg component had infolded.This diagnosis was confirmed on a cta performed on (b)(6) 2018.It was reported that the patient has some thrombus and claudication in the lower limbs.A reintervention is planned for (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative/corrected data: reintervention occurred on (b)(6) 2018, whereby two aortic extender components (pla280300) were implanted to treat the proximal infolding of the trunk-ipsilateral leg component.The patient tolerated the procedure.
|
| |
|