TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device is undergoing an evaluation but has not yet been completed.Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(4) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for lo display.The customer is concerned with results obtained of lo (true manger air app).The expected fasting blood glucose test result range is 100 -120mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is not stored by customer according to specification in the kitchen.The test strip lot manufacturer's expiration date is 07/17/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6) (not in the meter but on the true manger air app).
 
Event or Problem Description
Consumer reported complaint for lo display.The customer is concerned with results obtained of lo (true manger air app).The expected fasting blood glucose test result range is 100 -120mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is not stored by customer according to specification in the kitchen.The test strip lot manufacturer's expiration date is 07/17/2019 and open vial date is april 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly) result 1 :108 mg/dl date:(b)(6) 2018, time:06:23 am fasting, result 2 :100 mg/dl date:(b)(6) 2018, time:06:49 am fasting, result 3 :122 mg/dl date:(b)(6) 2018, time:08:35 am fasting, result 4 :lo date:(b)(6) 2018, time:n/a (not in the meter but on the true manger air app).
 
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report (b)(4).Returned meter evaluated with no defect found.Returned test strips evaluated with defect found.Qc investigation on (b)(6) 2018, returned test strips produce low readings.Most likely underlying root cause of low readings are due to improper storage: vial left open for an extended period of time.Test strip udi# (b)(4).Note: manufacturer contacted customer on 8/27/2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7877970
Report Number1000113657-2018-00951
Device Sequence Number10806536
Product Code NBW
UDI-Device Identifier00021292007584
UDI-Public(01)00021292007584
Combination Product (Y/N)N
Initial Reporter StateIL
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMV2750
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2018
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer08/21/2018
Initial Report FDA Received Date09/14/2018
Supplement Report FDA Received Date09/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY
-
-