Model Number TRUETRACK
Date of Event 08/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Product is working as designed.Customer disconnected the call.Most likely underlying root cause: mlc-62-user had poor technique.Test strip udi# (b)(4).Manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Unable to send product notification letter as no address on file.
 
Event or Problem Description
Consumer reported complaint for e-3 error code; used test strip, test strip outside of vial too long, sample on top of test strip.The e-3 error occurred when the blood sample was absorbed.Prior to call being transferred, a blood test had been performed by customer and produced test result of 441 mg/dl using truetrack meter; customer stated she was not concerned with the result.The customer's expected blood glucose test result range was not disclosed.At the time of the call, the customer stated that she was feeling funny and shaky.Medical attention was not needed at the time of the call.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 11/30/2020 and open vial date was undisclosed.The meter memory was not reviewed for previous test result history.Customer had disconnected the call during the troubleshooting process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7878363
Report Number1000113657-2018-00953
Device Sequence Number10811043
Product Code NBW
UDI-Device Identifier00011822577540
UDI-Public(01)00011822577540
Combination Product (Y/N)N
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/14/2018
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2020
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRV5270
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2018
Initial Report FDA Received Date09/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
-
-