• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF284
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary: the data files and balloon catheter, 2af284 with lot number 91562 were returned and analyzed.The data files showed at least 22 injections were performed on the date of the event and system notice # (b)(4) ¿the refrigerant delivery path is obstructed¿ and system notice # (b)(4) ¿the safety system has detected a compromised outer vacuum¿ were triggered.The first 7 applications were performed with the returned balloon catheter and the other 15 applications were performed with a different balloon catheter.Visual inspection of the returned catheter showed the device was intact with no apparent issues.Pressure test revealed a leak through the guide wire lumen.Dissection showed a guide wire lumen breach and kink at 1.4565 inches from the tip.In conclusion.The reported issues were confirm and the balloon catheter failed the returned product inspection due to a guide wire lumen breach and kink.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a cryo ablation procedure, a system notice was received indicating that the refrigerant delivery path was obstructed.The coaxial umbilical cable was replaced without resolve.The electrical cable and auto connection box were also replaced.A freeze was successful, however, a system notice was received indicating that the safety system detected a compromised outer vacuum.The balloon catheter was then replaced which resolved the issues.The case was completed with cryo.No patient complications have been reported as a result of this event.The balloon catheter was returned to the manufacturer, analyzed, and tested out of specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Common Device Name
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key7878788
Report Number3002648230-2018-00659
Device Sequence Number1985885
Product Code OAE
UDI-Device Identifier00643169753693
UDI-Public00643169753693
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P100010/S015
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/27/2019
Device Model Number2AF284
Device Catalogue Number2AF284
Device Lot Number91562
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2018
Initial Report FDA Received Date09/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight135
-
-