• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REF THREADED HOLE COVER; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REF THREADED HOLE COVER; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Catalog Number 71336500
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 02/16/2010
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a medical intervention was performed on (b)(6) 2010 due to swelling.Right hip bursa was injected.
 
Additional Manufacturer Narrative
The associated complaint devices was returned and evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A lab analysis indicated that the sleeve exhibits bone on/ingrowth on the medial side; the amount suggests the cone was well fixed prior to removal.The sleeve was well fixed is the missing porous coating on the proximal lateral area of the sleeve as this was likely the result of instrument contact.A potential fracture initiation site is visible at the lateral mating edge of the sleeve and stem component.A piece of the stem below the lower calcar platform was returned.An sem examination of the fracture surface of this piece supports the earlier observation of the potential crack initiation site being at the lateral mating surface of the sleeve and stem.Edxs analysis within the fracture site did not find any foreign or unexpected constituents in the material.The fracture surface reveals characteristics consistent with a fatigue fracture.A clinical analysis indicated that based on the available information, the root cause of the first revision was the broken stem that resulted from the stated fall.The clinical symptoms of the reported of "hip pain" and osteolysis of the greater trochanter may be consistent with an adverse reaction to metal debris but this cannot be confirmed based on the information provided as well this could also be lack of support around the proximal stem due to stress shielding.The root cause for the second revision due to "loosening" just 8 months later cannot be concluded.However, it is unknown if the use of third party components added to the issue of "loosening".No current information or patient history has been provided for this assessment.Should information become available this complaint can be re-assessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF THREADED HOLE COVER
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7879098
Report Number1020279-2018-01838
Device Sequence Number182216
Product Code MBL
UDI-Device Identifier03596010197634
UDI-Public03596010197634
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K070756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/11/2019
Device Catalogue Number71336500
Device Lot Number09DM08780
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2018
Initial Date Received by Manufacturer 08/16/2018
Supplement Date Received by Manufacturer10/16/2018
Initial Report FDA Received Date09/14/2018
Supplement Report FDA Received Date11/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age70 YR
-
-