PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Failure to Deliver Shock/Stimulation (1133)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted physio-control to report that their device would not shock during a cardiac arrest call.The customer advised that the device would not shock during two (2) attempts.Medical personnel had to call for a backup device.Once the backup device arrived, it was connected to the patient to continue care.There was no adverse effects to the patient as a result of the reported issue, as the backup device was used to continue patient care.
 
Additional Manufacturer Narrative
Physio-control evaluated the customer's device and was able to duplicate the reported issue.Physio determined that the cause of the reported issue was due to the therapy pcb assembly.Physio replaced the therapy pcb assembly to resolve the reported issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.Physio further evaluated the removed therapy pcb assembly and determined the cause of the reported issue was due to a transistor, designator q7 on the h-bridge that was shorted.This short caused the device to deliver monophasic shocks and intermittently would not shock when prompted.
 
Event or Problem Description
The customer contacted physio-control to report that their device would not shock during a cardiac arrest call.The customer advised that the device would not shock during two (2) attempts.Medical personnel had to call for a backup device.Once the backup device arrived, it was connected to the patient to continue care.There was no adverse effects to the patient as a result of the reported issue, as the backup device was used to continue patient care.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7879353
Report Number0003015876-2018-01533
Device Sequence Number2076995
Product Code MKJ
UDI-Device Identifier00883873871805
UDI-Public00883873871805
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001946
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2018
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer10/29/2018
Initial Report FDA Received Date09/14/2018
Supplement Report FDA Received Date11/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-