Model Number 88901
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Implant and explant dates: the implant or explant dates are not applicable to this device.
 
Event or Problem Description
This case was reviewed and investigated according to the manufacturer's policy.It was reported during a therapeutic peripheral procedure, there was no image and the screen was black when the manufacturer's device was in the patient's body.A second of the same make and model of the manufacturer's device was used to successfully complete the procedure.Visual and microscopic inspection were performed on the returned device.Image and functional testing could not be conducted as the pim connector was cut and not returned with the device.Additionally, there were scratches along the fillet and a small portion of the adhesive was missing.Based on the condition of the returned device, the probable cause of the reported failure (no image) could not be determined.The probable cause of the observed missing adhesive could not be conclusively determined by the investigation.Strain, impact, and forces associated with use can affect the integrity of the device.The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.This product problem is being reported because a portion of the device was missing and it could not be concluded when the separation occurred.There is a potential for harm if the malfunction were to recur.
 
Event or Problem Description
This case was reviewed and investigated according to the manufacturer's policy.Internal reference: (b)(4).This follow-up supplemental report #1 is being submitted to report additional investigation findings, to wit: there is no potential for harm if the malfunction were to recur.This complaint was reassessed based on an engineering study that indicated, this family of catheters do not exhibit any adhesive detachment during normal use.There is no evidence of abnormal use conditions, it is likely the adhesive detached post-procedure, during the process of returning the device for analysis.There was no patient injury, no status decline, no adverse event, and no unplanned additional medical or surgical intervention reported.There is no potential for harm if the malfunction were to recur.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to the alleged death or deterioration in the state of the health of any person.
 
Additional Manufacturer Narrative
H3 other text : placeholder.
 
Additional Manufacturer Narrative
Block h6: supplemental report #1 indicated that the adhesive likely detached post procedure; however, the conclusion code was not updated from 4315 (cause not established) to 4308 (cause traced to transport/storage).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONS PV .035 DIGITAL IVUS CATHETER
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
VOLCANO CORPORATION
2870 kilgore road
rancho cordova CA 95670
Manufacturer (Section G)
VOLCARICA S.R.L.
coyol free zone&business park
alajuela, B37
CS   B37
Manufacturer Contact
melissa pieplow
2870 kilgore road
rancho cordova, CA 95670
9163651925
MDR Report Key7879358
Report Number2939520-2018-00093
Device Sequence Number19345351
Product Code OBJ
UDI-Device Identifier00845225002329
UDI-Public(01)00845225002329(17)200131
Combination Product (Y/N)Y
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K153094
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/14/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Model Number88901
Device Catalogue Number400-0200.173
Device Lot Number0232 50128498
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/24/2018
Supplement Date Received by Manufacturer01/14/2005
12/13/2021
Initial Report FDA Received Date09/14/2018
Supplement Report FDA Received Date02/03/2020
12/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-