Model Number 15
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device but was unable to duplicate the reported issue.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted physio-control to report that their device would not complete the boot-up process in order to power on.The device will behave as though it's attempting to power on, but then immediately powers off.There was no patient use associated with the reported event.
 
Event or Problem Description
The customer contacted physio-control to report that their device would not complete the boot-up process in order to power on.The device will behave as though it's attempting to power on, but then immediately powers off.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Field g5 combination product of the initial mdr indicates ¿ blank field g5 combination product of the initial mdr should have indicated - no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7879820
Report Number0003015876-2018-01535
Device Sequence Number2508293
Product Code MKJ
UDI-Device Identifier00883873911457
UDI-Public00883873911457
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001268
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2018
Initial Date Received by Manufacturer 08/24/2018
Supplement Date Received by Manufacturer04/02/2020
Initial Report FDA Received Date09/16/2018
Supplement Report FDA Received Date04/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No