• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ONETOUCHPING GLUCOSEMGMTSYSTEM; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ONETOUCHPING GLUCOSEMGMTSYSTEM; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Hyperglycemia (1905); Vomiting (2144); Elevated ketones/Diabetic Ketoacidosis (2364)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.(b)(4).
 
Event or Problem Description
On (b)(6) 2018, a reporter contacted animas alleging that a patient had been hospitalized with hyperglycemia.The reporter stated that the patient had a blood glucose of 500 mg/dl with symptoms of large ketones, thirst, urination, and vomiting.The patient was receiving intravenous fluids and insulin via injections while at the hospital.The reporter stated that there was an issue with the pump¿s time and date settings causing the patient to receive programmed night time basal rates during the day and daytime basal rates at night.During troubleshooting it was determined that the pump was powering on to the verify screen for the date and time.This complaint is being reported based on the allegation that the patient experienced a hyperglycemic event as a result of the improper time being selected in the pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONETOUCHPING GLUCOSEMGMTSYSTEM
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key7880583
Report Number2531779-2018-16875
Device Sequence Number1939621
Product Code LZG
UDI-Device Identifier10840406100990
UDI-Public10840406100990
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K080639
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Device Age17 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2018
Initial Report FDA Received Date09/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age13 YR
-
-