C.R. BARD, INC. (GFO) LUTONIX 035 DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9004
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Reocclusion (1985)
Date of Event 08/17/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed there is no similar reocclusion complaints associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual samples were not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was possibly related to the study devices, but not related to the procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported through a post market clinical study that during the index procedure, two lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the right superficial femoral artery (sfa).Approximately 5 months after the index procedure, the patient¿s right sfa was reportedly reoccluded.It is unknown at this time, if a revascularization was performed.The investigator assessed that the event was possibly related to the study devices, but not related to the procedure.The samples were discarded by the user facility and are not available for evaluation.No adverse patient effects were reported.This is one of two products involved with the reported event and the associated manufacturer¿s report number is 3006513822-2018-00195.
 
Additional Manufacturer Narrative
Additional information: b.6 the relevant tests and lab data section had the following information added "on 17-aug-2018, dus imaging indicated target lesion was reoccluded." e.3 the complainant occupation - other had the following information added "cardiovascular internal medicine".Corrected data: g.1 the mdr contact person and email address were updated to "(b)(6)" and "(b)(6)".G.2 the phone number was updated to "(b)(6)".H3 analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed there is no similar reocclusion complaints associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual samples were not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was possibly related to the study devices, but not related to the procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through a post market clinical study that during the index procedure, two lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the right superficial femoral artery (sfa).Approximately 5 months after the index procedure, the patient¿s right sfa was reportedly reoccluded.It is unknown at this time, if a revascularization was performed.The investigator assessed that the event was possibly related to the study devices, but not related to the procedure.The samples were discarded by the user facility and are not available for evaluation.No adverse patient effects were reported.This is one of two products involved with the reported event and the associated manufacturer¿s report number is 3006513822-2018-00195.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
MDR Report Key7880780
Report Number3006513822-2018-00196
Device Sequence Number14040268
Product Code ONU
UDI-Device Identifier00801741126888
UDI-Public(01)00801741126888
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/13/2019
Device Model Number9004
Device Catalogue NumberLX351306805F
Device Lot NumberGFBT0672
Was Device Available for Evaluation? No
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer03/19/2019
Initial Report FDA Received Date09/17/2018
Supplement Report FDA Received Date03/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
Patient Weight57
-
-