| Catalog Number |
ADM06004013P |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the physician had used an in.Pact admiral device on a plaque lesion in the patients mid right superficial femoral artery but it had burst.The physician then took a second in.Pact admiral device with a 7f sheath and a 6mm spider fx embolic protection device for treatment of the lesion.Slight calcification and tortuosity were reported.The ifu was followed and the device was prepped, including drawing a negative pressure, without issue.An unknown inflation device was used.It was reported that pinhole balloon burst occurred on the device at an inflation pressure of 10atm on the first inflation of the device.The procedure was completed using plain old balloon angioplasty (poba).No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: the device was returned for evaluation.Visual and tactile inspections were performed on the device.The balloon was unfolded and dirty with blood.It was not possible to load the 0,035¿¿ guidewire due to the hardened blood inside the lumen.Negative pressure was applied to the balloon and the purging procedure failed.An attempt to inflate the balloon was made but a pinhole was detected and it was not possible to maintain the pressure.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|