| Model Number |
GL-323 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when they opened the suture intra-operatively, they noticed that the tip of the needle broke off from the suture.They noted that it was not used.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the opened samples noted one needle and one partial suture.The needle was received broken at the tip.Upon microscopic inspection, instrument marks were observed near the break site.As no sealed samples were returned, a bend moment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.The needle should always be held in the middle where the diameter is greatest.Grasping the needle near the swaged end or the tip could cause bends or breaks, or damage the connection between the needle and suture.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|