| Model Number |
72201919S |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the color went grey then the device blacked out.There was a surgical delayr greater than 30 minutes.The procedure was completed with a backup device with no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Ccu failed functional testing with blank video output.Cause of output malfunction is a defective electronic component on the video processor pcb.Product passed functional testing with a known good video processor pcb installed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
| |
|