• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SERVICE CCU HIGH DEF 560P 2NDGEN; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SERVICE CCU HIGH DEF 560P 2NDGEN; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 72201919S
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/24/2018
Event Type  malfunction  
Event Description
It was reported that the color went grey then the device blacked out.There was a surgical delayr greater than 30 minutes.The procedure was completed with a backup device with no patient injury reported.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Ccu failed functional testing with blank video output.Cause of output malfunction is a defective electronic component on the video processor pcb.Product passed functional testing with a known good video processor pcb installed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVICE CCU HIGH DEF 560P 2NDGEN
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key7881855
MDR Text Key120696528
Report Number1643264-2018-00719
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier03596010645517
UDI-Public(01)03596010645517
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070266
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72201919S
Device Catalogue Number72201919S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/15/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-