| Model Number |
72201919F |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the color went grey then the device blacked out.There was a surgical delay greater than 30 minutes.The procedure was completed with a backup device with no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Ccu failed functional testing with no live video output.Cause of video output malfunction is a defective fan on the analog output pcb.The defective fan resulted in the video processor pcb overheating and live video output malfunction.The complaint was confirmed and a root cause has been associated with an electrical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
| |
|