Model Number 72201919F
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/25/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the color went grey then the device blacked out.There was a surgical delay greater than 30 minutes.The procedure was completed with a backup device with no patient injury reported.
 
Additional Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Ccu failed functional testing with no live video output.Cause of video output malfunction is a defective fan on the analog output pcb.The defective fan resulted in the video processor pcb overheating and live video output malfunction.The complaint was confirmed and a root cause has been associated with an electrical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFB CCU HIGH DEF 560P 2NDGEN
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7881867
Report Number1643264-2018-00718
Device Sequence Number19348150
Product Code GCJ
UDI-Device Identifier00885556572818
UDI-Public(01)00885556572818
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K070266
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72201919F
Device Catalogue Number72201919F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Initial Date Received by Manufacturer 08/27/2018
Supplement Date Received by Manufacturer10/15/2018
Initial Report FDA Received Date09/17/2018
Supplement Report FDA Received Date10/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-