| Catalog Number |
225361 |
| Medical Device Problem Code |
Melted (1385)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4)-incomplete.The expiration date is not currently available.
|
| |
|
Event or Problem Description
|
|
It was reported that the probe tip was deformed like melted during the surgery on (b)(6) 2018.Then, the surgeon stopped using the reported probe.The surgery was completed by using alternative probe (same p/n, lot number was unknown).It was brand new and the first use when the issue occurred.There was no surgical delay and no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received and evaluated.Visual observation of the active tip indicates signs of activation and melted manifold between 6 - 9 o'clock position, confirming this complaint.No functional test was conducted to avoid further damage to the device and to test equipment.A supplier corrective and preventative action (capa) was completed to address this issue.The root cause from the investigation is attributed to the design which requires that the operator should have a specific aptitude to ensure the glue wicks successfully; this aptitude is not presently ensured by the training program or by the electrode design.The electrode design relies on the epotek adhesive to provide a dielectric barrier between active and return path.The internal tracking distance is shorter than the external; therefore if the epotek layer breaks down, internal activation can occur with resulting sparking and arching.As identified by previous supplier capa, the occurrence rate of the capa failure mode can be reduced to an acceptable level by increasing the aptitude of the operator and this can be further improved by changing the training requirement when assembly aids are updated to a repeat training requirement on a six (6) month basis.A device history record (dhr) review has been conducted and the results indicate that this batch of product was processed without any incident and therefore there is no evidence of manufacturing anomalies on the paperwork reviewed.Depuy mitek has confirmed the date of manufacture of this lot was prior to the training requirement change from the supplier capa.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
|
| |
|