| Model Number |
MMT-723LNAP |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone that the insulin pump had button error alarm.Customer¿s blood glucose was unknown at the time of incident.Customer stated that the blank display as well as continuous noise.Insulin pump was exposed to moisture.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with blank display and constant tone due to moisture damage on the electronics assembly.Unable to perform all functional testing including the operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime and excessive no delivery test or verify button keypad anomaly due to blank display.Insulin pump received with moisture damage motor, vibrator, keypad and battery tube assembly.
|
| |
|