U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ULTRA FAST-FIX ASSEMBLY - CURVED; STAPLE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ULTRA FAST-FIX ASSEMBLY - CURVED; STAPLE, FIXATION, BONE Back to Search Results
Model Number 72201491
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/10/2018
Event Type  malfunction  
Event Description
It was reported that during the meniscus repair surgery,after opening the package and cutting out the limiting deep sleeve, surgeon found that there was a protrusion of the suture between the two ts.As a result, the second t was also taken out after the first needle was inserted.There was a surgical delay greater than 30 minutes.
 
Manufacturer Narrative
One curved ultra fast-fix assembly was returned for evaluation.Visual assessment showed the depth straw has been trimmed to the surgeons desired length.No ts or suture remain on the delivery needle but were returned.Examination of the construct showed it has been cinched, no abnormalities were observed with the suture or ts.This investigation could not identify any evidence of product contribution to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA FAST-FIX ASSEMBLY - CURVED
Type of Device
STAPLE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key7882982
MDR Text Key120554775
Report Number1219602-2018-01261
Device Sequence Number1
Product Code JDR
UDI-Device Identifier03596010597205
UDI-Public(01)03596010597205(17)211219(10)50637332
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K972326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/19/2021
Device Model Number72201491
Device Catalogue Number72201491
Device Lot Number50637332
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2018
Date Manufacturer Received10/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age27 YR
Patient Weight65
-
-