| Catalog Number |
6260-9-144 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
08/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported relevant discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before (b)(6) 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.Not returned.
|
| |
|
Event or Problem Description
|
|
Surgeon revised a stem, head and liner due to trunnionosis.Original surgery was (b)(6) 2007.Update 23/august/2018: rep provided explant pictures, pre-revision x-rays, primary operative report, and usage sheets from primary and revision procedures.Explant picture shows wear on the rim of the poly insert and pre-revision x-rays show the head disassociated from the stem.
|
| |
|
Event or Problem Description
|
|
Surgeon revised a stem, head and liner due to trunnionosis.Original surgery was (b)(6) 2007.Update (b)(6) 2018: rep provided explant pictures, pre-revision x-rays, primary operative report, and usage sheets from primary and revision procedures.Explant picture shows wear on the rim of the poly insert and pre-revision x-rays show the head disassociated from the stem.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding disassociation involving an accolade stem was reported.The event was confirmed on review of the provided medical records by a clinical consultant.Method & results: device evaluation and results: a review of the provided photo of the explanted head showed nothing remarkable.Medical records received and evaluation a review of the provided medical records by a clinical physician noted; review of these records confirms femoral head dissociation from a tha femoral stem occurred secondary to remodeling and loss of material of the femoral stem trunnion.The root cause cannot be determined as insufficient information was available device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: review indicated there have been no other similar events for the reported lot.Conclusion: the event was confirmed on review of the provided medical records by a clinical physician however the root cause could not be determined.It is noted that the subject device has been identified to be within scope of an nc and capa.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of the nc and capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before (b)(6) 2011.The root cause analyses identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.No further investigation is required.
|
| |
|