• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 44MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 44MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 623-00-44G
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code Injury (2348)
Date of Event 08/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported relevant discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Surgeon revised a stem, head, and liner due to trunnionosis.Original surgery was (b)(6) 2007.Update 23/august/2018: rep provided explant pictures, pre-revision x-rays, primary operative report, and usage sheets from primary and revision procedures.Explant picture shows wear on the rim of the poly insert and pre-revision x-rays show the head disassociated from the stem.
 
Event or Problem Description
Surgeon revised a stem, head, and liner due to trunnionosis.Original surgery was (b)(6) 2007.Update 23/august/2018: rep provided explant pictures, pre-revision x-rays, primary operative report, and usage sheets from primary and revision procedures.Explant picture shows wear on the rim of the poly insert and pre-revision x-rays show the head disassociated from the stem.
 
Additional Manufacturer Narrative
An event regarding wear involving a trident liner was reported.The event was not confirmed.Visual inspection: although the device was not returned an image of the explanted liner was made available.The image shows damage to the rim of the liner.It cannot be confirmed if this damage is due to wear or the explantation process.A review of the provided medical records and x-rays by a clinical consultant indicated: review of these records confirms femoral head dissociation from a tha femoral stem occurred secondary to remodeling and loss of material of the femoral stem trunnion.The root cause cannot be determined as insufficient information was available documentation which would aid in the further completion of this assessment would include: material analysis of the explanted implants.All devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the reported lot.The event and exact cause of the event could not be confirmed or determined because insufficient information was provided.Additional information, including the return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG INSERT 44MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7883143
Report Number0002249697-2018-02910
Device Sequence Number2725772
Product Code LPH
UDI-Device Identifier07613327025385
UDI-Public07613327025385
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K062419
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2012
Device Catalogue Number623-00-44G
Device Lot NumberLE3MRD
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer10/18/2018
Initial Report FDA Received Date09/17/2018
Supplement Report FDA Received Date11/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age76 YR
-
-