| Catalog Number |
140.640 |
| Medical Device Problem Codes |
Failure to Osseointegrate (1863); Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Tissue Damage (2104)
|
| Date of Event |
08/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reported that nearly 1 year after the dental implant and abutment were placed in ada site 23 in the mouth, the implant lost osseointegration in type ii bone.Clinician reports patient's pain.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.After the evaluation, the event was considered as osseointegration failure.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.Moreover, it was seen that the dentist has used more drills than recommended to perform the implant installation.The dentist reported that immediate load and immediate implant procedures were performed.
|
| |
|
Event or Problem Description
|
|
The clinician reported that nearly 1 year after the dental implant and abutment were placed in ada site 23 in the mouth, the implant lost osseointegration in type ii bone.Clinician reports patient's pain.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.(b)(4).
|
| |
|