| Model Number |
ZCT150 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
08/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zct150 17.5 diopter intraocular lens (iol) was implanted in the patient's right eye (od) on (b)(6) 2018.It was later explanted due to a hyperopic outcome on (b)(6) 2018.The replacement lens was a zct150, a larger 18.5 diopter lens.There was no incision enlargement, vitrectomy, or sutures used.Reportedly, the patient is doing well.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 10/1/2018.Device returned to manufacturer? yes.Device evaluation: visual inspection shows the product is cut into two pieces, most probably to make explant possible.Considering the condition of the lens, additional analysis is not possible.The complaint cannot be confirmed.Manufacturing record review: during the manufacturing record review of the production order for this serial number, no product deficiency was identified.The documentation shows that the production order was manufactured according to specifications.Labeling review: the labeling review was completed and revealed that the directions for use (dfu) provide the customer with proper usage instructions and guidelines.Complaint history: a search of complaints revealed that no similar complaints were received for this production order number.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|