| Model Number |
31G X 5MM |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This defect was most probably caused by the higher than usual force holding the cannula inside the rubber plug of the vial, exceeding the strength of bond of cannula and needle hub.
|
| |
|
Event or Problem Description
|
|
Patient is having issues removing the pen needle from her insulin pen.Customer has not been injured.
|
| |
|
Search Alerts/Recalls
|